Rotors and accessories
26
Closing the aerosol-tight bio-containment
After greasing the seal, turn the cover until it sits lightly
on the bucket.
To achieve uniform seal, turn the cover clockwise by
1½ grip area ( approx. 15°). Use the setting marks on
the bucket as a guide.
Micro Liter Rotor 24 x 2.0 ml
Attention :
Please check that your sample containers are suitable
for the centrifugal application desired.
Please observe the permissible filling volumes!
Nominal volume:
Permissible volume:
2.0
ml
-
1.5 ml
others
-
2
/
3
nominal volume
Bucket mark
Grip area
Содержание accuSpin 3
Страница 1: ...accuSpin 3 3R OPERATING INSTRUCTIONS...
Страница 10: ...For your safety 6 Notes...
Страница 16: ...Description of the accuSpin 12 Notes...
Страница 20: ...Rotors and accessories 16...
Страница 32: ...Rotors and accessories 28 Notes...
Страница 38: ...Operation 34...
Страница 68: ...64 Notes...
Страница 70: ...66 Acceleration profiles Swinging bucket rotor with rectangular buckets 7500 4395 time s speed rpm 9 1...
Страница 71: ...67 speed rpm time s 1 9 Deceleration profiles Swinging bucket rotor with rectangular buckets 7500 4395...
Страница 72: ...68 Acceleration profiles Micro liter Rotor 7500 3337 speed rpm time s...
Страница 73: ...69 Deceleration profiles Micro liter Rotor 7500 3337 speed rpm time s 9 1...
Страница 76: ...72...