Operation
44
Audible alarm
Accompanying all error messages, a warning signal is
given out which only is silenced upon pressing any key.
As an option, you can also have the end of a run sig-
naled acoustically. To activate or deactivate this option
press the program key
and switch on the centrifuge
simultaneously. As far as you keep this key pressed,
the selection mode is active.
The speed panel shows "beep“ and the time panel
shows "on“ or "off“.
By actuating the upward-key in the time panel the
signal function can be switched on or off
.
When the message "rotor“ flashes, pressing
the start key once is sufficient to turn off the
warning signal and to accelerate the rotor to
the maximum speed displayed by the in-
strument.
Turning the centrifuge off
By switching the main switch into "0“ position the cen-
trifuge is turned off.
The main power switch should be turned off
after a complete centrifugation run. Without
motor deceleration, it takes much more time
until the rotor comes to a halt.
The centrifuge lid can only be opened automatically if
the centrifuge is turned on!
Содержание accuSpin 3
Страница 1: ...accuSpin 3 3R OPERATING INSTRUCTIONS...
Страница 10: ...For your safety 6 Notes...
Страница 16: ...Description of the accuSpin 12 Notes...
Страница 20: ...Rotors and accessories 16...
Страница 32: ...Rotors and accessories 28 Notes...
Страница 38: ...Operation 34...
Страница 68: ...64 Notes...
Страница 70: ...66 Acceleration profiles Swinging bucket rotor with rectangular buckets 7500 4395 time s speed rpm 9 1...
Страница 71: ...67 speed rpm time s 1 9 Deceleration profiles Swinging bucket rotor with rectangular buckets 7500 4395...
Страница 72: ...68 Acceleration profiles Micro liter Rotor 7500 3337 speed rpm time s...
Страница 73: ...69 Deceleration profiles Micro liter Rotor 7500 3337 speed rpm time s 9 1...
Страница 76: ...72...