Troubleshooting
49
Troubleshooting
Emergency lid release
In case of a power failure the lid could not be opened
normally using the electrical lid unlocking mechanism
.
To permit unloading in this case, the centrifuge is
equipped with an emergency override release. How-
ever, you may use this system in case of emergency
only.
Rotors can spin at high speed! Touch-
ing it may cause severe injuries!
Always wait for several minutes until
the rotor has come to a complete stop.
Without power the brake does not
function, and deceleration takes much
longer than normal!
Proceed as follows:
1. Make sure that the rotor is at a stand still
(observe through window in the cover).
During a power failure it is impossible
to lock the lid once the emergency lid
release has been used. Never stop the
rotor using your hands or tools!
2. Unplug the main power cord.
3. Facing the centrifuge under the left-hand side there
is a plastic plug that can be removed using a
screwdriver. After removal, use your fingers or a
pair of pliers to pull the attached cord and activate
the manual door override. The lid will open and you
can remove your samples.
Emergency release
Содержание accuSpin 3
Страница 1: ...accuSpin 3 3R OPERATING INSTRUCTIONS...
Страница 10: ...For your safety 6 Notes...
Страница 16: ...Description of the accuSpin 12 Notes...
Страница 20: ...Rotors and accessories 16...
Страница 32: ...Rotors and accessories 28 Notes...
Страница 38: ...Operation 34...
Страница 68: ...64 Notes...
Страница 70: ...66 Acceleration profiles Swinging bucket rotor with rectangular buckets 7500 4395 time s speed rpm 9 1...
Страница 71: ...67 speed rpm time s 1 9 Deceleration profiles Swinging bucket rotor with rectangular buckets 7500 4395...
Страница 72: ...68 Acceleration profiles Micro liter Rotor 7500 3337 speed rpm time s...
Страница 73: ...69 Deceleration profiles Micro liter Rotor 7500 3337 speed rpm time s 9 1...
Страница 76: ...72...