Operation
41
Stopping the centrifuge
Stopping with preset run time
Normally the run time has been selected, and all you
have to do is wait until the centrifuge terminates the run
automatically at the end of the set time.
As soon as the speed reaches zero, the display reads
"End". You can now open the centrifuge by pressing
the "open lid" key
and remove your samples.
If the lid has not been fully lifted out of the lid lock, the
message "Lift lid" appears (you must manually lift the
lid).
You can manually stop the centrifuge at any time by
pressing the "Stop“ key
.
At this point the remaining run time is displayed.
Stopping with continuous operation
If you have chosen continuous operation, you must
stop the centrifuge manually. Press the "Stop" key
in the control panel. The centrifuge starts deceleration
with the preset deceleration profile.
The display reads "End", and you can open the lid by
pressing the "Open lid" key
and remove your sam-
ples.
Temperature control during standby
Temperature control becomes active once the rotor
has been identified. This is the case after a centrifuga-
tion run exceeding 300 rpm. At standby the display
reads "End".
If the rotor has not been identified (lid has been closed
and the "Start" key has not yet been pressed, speed
panel shows "0" with flashing point), the instrument
regulates the temperature so that the samples cannot
freeze in any one of the usable rotors.
If you find the systematic deviation of up to 4 K not
being sufficient, you have to start the rotor for a short
period until it is identified.
Содержание accuSpin 3
Страница 1: ...accuSpin 3 3R OPERATING INSTRUCTIONS...
Страница 10: ...For your safety 6 Notes...
Страница 16: ...Description of the accuSpin 12 Notes...
Страница 20: ...Rotors and accessories 16...
Страница 32: ...Rotors and accessories 28 Notes...
Страница 38: ...Operation 34...
Страница 68: ...64 Notes...
Страница 70: ...66 Acceleration profiles Swinging bucket rotor with rectangular buckets 7500 4395 time s speed rpm 9 1...
Страница 71: ...67 speed rpm time s 1 9 Deceleration profiles Swinging bucket rotor with rectangular buckets 7500 4395...
Страница 72: ...68 Acceleration profiles Micro liter Rotor 7500 3337 speed rpm time s...
Страница 73: ...69 Deceleration profiles Micro liter Rotor 7500 3337 speed rpm time s 9 1...
Страница 76: ...72...