Technical Data
63
Electrical connections / fuses
Order no.
Voltage
Frequency Nominal
current
Power
consumption
Fuse protection of
instrument
-
safety fuse
* thermic excess current
release
Fuse
protection of
building
accuSpin
3
7500 4391
120 V
60 Hz
10,2 A
900 W
15 A
15 AT
accuSpin
3
7500 4386
230 V
50/60 Hz
8.6 A
1250 W
10 A
16 AT
accuSpin
3R
7500 4392
120 V
60 Hz
12,0 A
1220 W
15 A
*
16
A
15 AT
accuSpin 3R
7500 4387
230 V
50/60 Hz
10.6 A
1900 W
10 A
*
16
A
16 AT