Troubleshooting
56
Error
Symptom
Possible causes and corrective measures
E-21
Instrument does not start or
brakes to standstill.
No rotor present or rotor identification impossible
A) Check whether a certified rotor is inserted.
B) Following a brief power failure, the rotor could not be
identified. Switch the instrument off and on again using
the main switch.
E-22
|
E-23
Rotor stops without brake
to standstill.
Instrument cannot be
operated.
Error in speed measurement
Switch the instrument off and on again. If the error persists,
contact service
E-24
|
E-27
Instrument cannot be
operated.
Wrong status information from lid lock.
Switch the instrument off and on again. If the error persists,
contact service.
E-28
Instrument decelerates
after a short starting-up.
Motor runs in wrong rotational direction.
Switch the instrument off and on again. If the error persists,
contact service.
Содержание accuSpin 3
Страница 1: ...accuSpin 3 3R OPERATING INSTRUCTIONS...
Страница 10: ...For your safety 6 Notes...
Страница 16: ...Description of the accuSpin 12 Notes...
Страница 20: ...Rotors and accessories 16...
Страница 32: ...Rotors and accessories 28 Notes...
Страница 38: ...Operation 34...
Страница 68: ...64 Notes...
Страница 70: ...66 Acceleration profiles Swinging bucket rotor with rectangular buckets 7500 4395 time s speed rpm 9 1...
Страница 71: ...67 speed rpm time s 1 9 Deceleration profiles Swinging bucket rotor with rectangular buckets 7500 4395...
Страница 72: ...68 Acceleration profiles Micro liter Rotor 7500 3337 speed rpm time s...
Страница 73: ...69 Deceleration profiles Micro liter Rotor 7500 3337 speed rpm time s 9 1...
Страница 76: ...72...