Troubleshooting
59
Error
Symptom
Possible causes and corrective measures
E-40
Rotor stops without brake
to standstill.
Instrument cannot be
operated.
Acceleration of rotor too slow
Open the lid and check the load!
-
Significant imbalance caused by missing bucket
-
Tubes or adapter do not fit
(rotor is rubbing the chamber or the cover)
-
Micro plates carriers on the motor cover
If the error persists, contact service.