Maintenance and care
46
During cleaning liquids and espe-
cially organic solvents should not
come into contact with the drive
shaft and the ball bearing.
Organic solvents may decompose
the lubricant of the motor bearing.
The drive shaft may block.
Instruments with refrigeration unit:
If a strong ice sheet is present in the internal
chamber, be sure to remove all condensate
after defrosting!
Please control and clean the venting slots
regularly!
Before cleaning the venting slots
please disconnect the centrifuge
from the mains supply.
Please pull mains plug!
Disinfection
If a centrifuge tube containing infectious material leaks
during a run, you have to disinfect the centrifuge im-
mediately.
Infectious material could enter the centri-
fuge if spills or tube breakage occur
.
Danger of infection may occur upon
contact
!
Take appropriate protective meas-
ures for personnel!
Mind the permissible filling volumes and
loading limits for the tubes
!
In case of contamination the operator has
to make sure, that no further persons are
jeopardized
!
Contaminated parts have to be decontami-
nated immediately
.
If required further protective measures
have to be initiated
.
Содержание accuSpin 3
Страница 1: ...accuSpin 3 3R OPERATING INSTRUCTIONS...
Страница 10: ...For your safety 6 Notes...
Страница 16: ...Description of the accuSpin 12 Notes...
Страница 20: ...Rotors and accessories 16...
Страница 32: ...Rotors and accessories 28 Notes...
Страница 38: ...Operation 34...
Страница 68: ...64 Notes...
Страница 70: ...66 Acceleration profiles Swinging bucket rotor with rectangular buckets 7500 4395 time s speed rpm 9 1...
Страница 71: ...67 speed rpm time s 1 9 Deceleration profiles Swinging bucket rotor with rectangular buckets 7500 4395...
Страница 72: ...68 Acceleration profiles Micro liter Rotor 7500 3337 speed rpm time s...
Страница 73: ...69 Deceleration profiles Micro liter Rotor 7500 3337 speed rpm time s 9 1...
Страница 76: ...72...