M a i n t e n a n c e
55
Model 1010 Ion Mill
Standard Magnification Version
Remove rotator
1.
Remove the two screws located within the plastic washers that are positioned at the
chamber-rotator plate interface.
2.
Firmly grasp the rotator assembly and pull toward the rear of the instrument.
It may be necessary to rotate the assembly so that the lower gear clears the motor
mounts.
CAUTION: Never set the stage so that it is being supported by the rotator
plate and the end of the specimen stage. Damage to the specimen
stage may result.
3.
Take the specimen stage rotator assembly to a clean work area for inspection,
cleaning, and any necessary repairs.
CAUTION: Extreme care must be taken when handling or conducting
service on the rotator plate. Any scratches or nicks on the face of
the rotator plate will lead to vacuum leaks.
Install rotator
1.
Prior to installation, remove the o-ring from the chamber and clean thoroughly with
methanol or isopropyl alcohol.
2.
Wipe the chamber surface with acetone.
3.
Allow to dry.
4.
Apply a light film of Fomblin® grease to the o-ring and position the o-ring into the
chamber.
5.
Install the rotator assembly into the chamber.
Make sure that the o-ring does not slip out of its groove during the installation of the
rotator.
There is a precise fit between the chamber and the bearing and the bearing and the
outer diameter of the rotator plate. The rotator plate should be installed as straight as
possible. Any angling of the rotator plate in relation to the chamber will prevent full
insertion of the rotator plate leading to the inability to achieve a vacuum.
Screws in plastic washers