O p e r a t i o n
26
Model 1010 Ion Mill
Standard Magnification Version
8.
To create a log file, click
Yes
.
Note: Creating a log file is not required for performing a milling operation.
9.
Provide a file name.
The ion milling process will begin. The display area of the ion mill program window
shows the progress of the operation, including stage temperature, rocking angle, ion
source voltage and current, milling angle, and status of shutter, lights, and chamber
insert.
Milling operation underway
The Start button changes to a “Stop” button, which allows the user to manually
terminate the milling operation at any time.
At the end of the ion milling program, the system initiates a resetting chamber
sequence, which includes purging the gas line, returning the stage to horizontal, and
recentering it. An audible signal notifies the user that the resetting chamber sequence
is complete.
10.
Click
OK
.
Click to Stop.