M a i n t e n a n c e
34
Model 1010 Ion Mill
Standard Magnification Version
Specimen chamber window
Caution: This window is located on the top chamber surface and on the specimen
axis of rotation, and requires periodic cleaning. Failure to properly maintain this
window will result in diminished sensitivity of the laser detector.
1.
Ensure the instrument is vented and not powered ON.
2.
To remove, firmly grasp the knurled portion of the bezel, then rotate
counterclockwise. The sapphire window is permanently adhered to the bezel to
assist is handling. Attempting to remove the window from the bezel will result in
damage to both components. Should the window become broken, return the
bezel/window to E.A. Fischione Instruments for repair.
3.
Remove sputtered material from the window surface using any type of metal or
glass polish.
4.
Use ultrasonic cleaning for final cleaning.
Top light window
1.
Ensure that the instrument is vented and not powered ON.
2.
Remove the 8-32 socket head cap screw and angled bracket that press the top
light assembly into its receptacle
3.
Open the front chamber access door
4.
Placing a gloved finger within the specimen chamber, gently press upward on the
top light assembly until it exits its receptacle.
5.
Locate the window shield, and engage the pins the wrench with the holes in the
shield.
6.
Rotate the wrench counterclockwise until the shield is removed.
Specimen chamber
window