M a i n t e n a n c e
39
Model 1010 Ion Mill
Standard Magnification Version
WARNING! ELECTRICAL HAZARD. High voltages can cause severe injury or
death. Turn off the instrument. Wait one minute to ensure that any
charged electronic component will discharge before disconnecting
the power cord from the instrument.
8.
Disconnect the power cord.
Disconnect gas line
1.
Disconnect the gas inlet to the ion source by firmly grasping the hexagonal nut that
connects the gas line to the cathode and rotating it counterclockwise.
A stainless steel ferrule on the stainless steel gas tube maintains the vacuum integrity
between of the gas tube and the fitting.
Allow the stainless steel ferrule to remain on the stainless steel gas tube.
Remove ion source cartridge
Note: The ion source functional components are all contained in the ion source
cartridge.
CAUTION: Ion source components are easily damaged. Work in a clean area.
The individual doing the work should be knowledgeable in the
polishing and assembly of precision mechanical components.
1.
Remove the three socket screws that attach the cartridge to the flange.
Note that the ion source ground wire (green) is connected to one of the screws.
Hexagonal nut
Stainless steel tube