M a i n t e n a n c e
31
Model 1010 Ion Mill
Standard Magnification Version
4.
Wash the filters in warm, soapy water then let air dry.
5.
Reinstall the filters by snapping them onto their respective mounts.
Changing the Fuses
WARNING! ELECTRICAL HAZARD. High voltages can cause severe injury or
death. Turn off the instrument. Wait one minute to ensure that any
charged electronic component will discharge before disconnecting
the power cord from the instrument.
1.
Turn the Model 1010 power switch to the off position.
2.
Disconnect the power cord.
3.
Using a small-blade screwdriver, gently remove the plastic cover from the power
entry module on the back of the instrument.
4.
Pull the fuse cartridge out of the power entry module.
5.
Replace the fuses.
For 110 VAC operation, replace two 6-amp fuses.
For 220 VAC operation, replace two 3-amp fuses.
Filter
Filter cover
Fuse cartridge
Fuse