FEASA LED ANALYSER
FUNCTIONAL VERSION
Sequence Capture Mode
GetFrequency## -
Returns the frequency and duty cycle of Led
Transmit
Receive
getfrequency##
Fff.f Hz xxx.x %
Where:
##
represents the Fiber Number and is a value in the range 01 - 20.
fff.f Hz
represents the Frequency in Hertz.
Xxx.x %
represents the Duty Cycle in Percentage.
Description
This command reports the Frequency in Hertz and the Duty Cycle in % of the Led under test
after the capture is complete.
Example:
The PC transmits
getfrequency01
to the LED Analyser to instruct it to report the Frequency
and duty cycle of fiber 1
getfrequency01
123.4Hz 067.8%