FEASA LED ANALYSER
FUNCTIONAL VERSION
Sequence Capture Mode
GetSequenceTimesAll
- Get the Duty Cycle of the Sequence for all
fibers
Transmit
Receive
getsequencetimesall
Toff1 Ton Toff2
Where:
ALL
represents the total number of Fibers used in the sequence capture.
Toff1 represents the time the LED is off at the beginning of the sequence.
Ton represents the time the LED is on during the sequence.
Toff2 represents the time the LED is off at the end of the sequence.
Description
This command returns three time intervals for each Fiber. The times are in milliseconds and
represent the time the LED is 'off' at the beginning, the 'on' time and the 'off' time at the end
of the sequence. A valid
capturesequence
command must be issued before this command in
order to get valid data. The resolution of the times Toff1, Ton and Toff2 depends on the
sequence capture and sequence wait time settings.
Example:
The PC transmits
getsequencetimesall
to the LED Analyser to instruct it to output the time
intervals for Fiber 01, Fiber 02, Fiber 03, Fiber 04, ......
getsequencetimesall
Toff1 = 00350ms Ton = 00340 ms Toff2 = 76910ms
Toff1 = 00350ms Ton = 00340 ms Toff2 = 76910ms
Toff1 = 00350ms Ton = 00340 ms Toff2 = 76910ms
Toff1 = 00350ms Ton = 00340 ms Toff2 = 76910ms
.
.
.
.
92
Toff1
Ton
Toff2
1
0
Содержание 20-F
Страница 11: ...FEASA LED ANALYSER FUNCTIONAL VERSION Figure 3b Intensity for the LED Analyser in LINEAR Mode Back to Index 11...
Страница 126: ...FEASA LED ANALYSER FUNCTIONAL VERSION User Software Back to Index 126...
Страница 127: ...FEASA LED ANALYSER FUNCTIONAL VERSION Test Software Back to Index 127...
Страница 128: ...FEASA LED ANALYSER FUNCTIONAL VERSION User Calibration Back to Index 128...
Страница 129: ...FEASA LED ANALYSER FUNCTIONAL VERSION Programming Back to Index 129...