FEASA LED ANALYSER
FUNCTIONAL VERSION
Sequence Capture Mode
Set/Put SequenceChannel@@ -
Select sequence mode Fiber No.
Transmit
Receive
setsequencechannel@@
putsequencechannel@@
OK
OK
Where:
@@
represents the Fiber and is a number in the range 01 -20.
Description
This command is used to set the Fiber number associated with the
Capture Sequence
command. The command
putSequencemode1
should be executed before this command.
In addition, the commands
putsequencecapture@@
,
putsequencewait##
and
putsequencenumber###
should be issued before this command.
The Analyser can store up to 9999 captures in memory when this mode is selected.
putsequencechannel@@
commands are stored in RAM memory while
setsequencechannel@@
commands are stored in Flash memory.
Example:
The PC transmits
putsequencechannel04
to the LED Analyser to instruct it to do sequence
capture on Fiber 04 only. The LED Analyser will return a string
OK
to the PC.
putsequencechannel04
OK
To determine which Fiber Number is set, use the command
getsequencechannel
.
It will
return a number in the range 01 to 20.
getsequencechannel
05
(Fiber Number 5)
79
Содержание 20-F
Страница 11: ...FEASA LED ANALYSER FUNCTIONAL VERSION Figure 3b Intensity for the LED Analyser in LINEAR Mode Back to Index 11...
Страница 126: ...FEASA LED ANALYSER FUNCTIONAL VERSION User Software Back to Index 126...
Страница 127: ...FEASA LED ANALYSER FUNCTIONAL VERSION Test Software Back to Index 127...
Страница 128: ...FEASA LED ANALYSER FUNCTIONAL VERSION User Calibration Back to Index 128...
Страница 129: ...FEASA LED ANALYSER FUNCTIONAL VERSION Programming Back to Index 129...