FEASA LED ANALYSER
FUNCTIONAL VERSION
Sequence Capture Mode
Set/Put SequenceNumber###(#) -
Set the Number of Captures
Transmit
Receive
putsequencenumber###
setsequencenumber###
putsequencenumber####
setsequencenumber####
OK
OK
OK
OK
Where:
###
represents the number of captures to store in the range 000 - 999.
####
represents the number of captures to store in the range 000 – 9999.
Description
This command is used to set the number of captures or samples to store. If the
se
tsequencemode# or putsequencemode#
is set to 0 then the number of samples that
can be stored is 3500.
If the
se
tsequencemode# or putsequencemode#
is set to 1 then the number of stored
samples increases to 9999.
PutSequencenumber###, PutSequencenumber####
commands are stored in RAM
memory while
S
etSequencenumber###,
S
etSequencenumber####
commands are
stored in Flash memory.
Example:
The PC transmits
putsequencenumber050
to the LED Analyser to instruct it to store 50
captures when the
CaptureSequence
command is issued. The LED Analyser will return a
string
OK
to the PC.
putsequencenumber050
OK
To determine the number of Sequence Captures stored, use the command
getsequencenumber
It will return a number in the range 0 - 9999
getsequencenumber
0250
(250 Captures)
82
Содержание 20-F
Страница 11: ...FEASA LED ANALYSER FUNCTIONAL VERSION Figure 3b Intensity for the LED Analyser in LINEAR Mode Back to Index 11...
Страница 126: ...FEASA LED ANALYSER FUNCTIONAL VERSION User Software Back to Index 126...
Страница 127: ...FEASA LED ANALYSER FUNCTIONAL VERSION Test Software Back to Index 127...
Страница 128: ...FEASA LED ANALYSER FUNCTIONAL VERSION User Calibration Back to Index 128...
Страница 129: ...FEASA LED ANALYSER FUNCTIONAL VERSION Programming Back to Index 129...