A - 28 |
Airvo 3
- USER MANUAL
English
7.4 Viewing alarm details
Action buttons
Hide suggestions
Mute alarm
Mute
Hide suggestions
Suggestions/Action
Alarm condition
Alarm condition
Show suggestions
Reset & dismiss
Mute alarm
Unmute alarm
Action buttons
Reset & dismiss
Unmute alarm
Hide
suggestions
Suggestions/
action
Audio pause
Alarm condition
Alarm condition
Alarms are displayed with suggestions and action buttons
for managing the alarm information:
• Tap the Audio Pause button to silence the alarm for
120 seconds. The Audio Pause button will change to
when audible alarms are silenced.
• Use
to scroll through multiple suggestions.
Some alarms have only one suggested resolution.
• Tap Hide suggestions to collapse the alarm
information to the Message bar. Restore
suggestions by tapping the alarm condition
on the Message bar.
The alarm condition and action button are displayed
on the Message bar when alarm information is
collapsed.
The Message bar cycles through each alarm condition
when multiple alarms are active. Tapping the Message
bar displays a list of active alarm conditions, from
highest priority to lowest priority and they are ordered
from when they occurred.
Alarm signals always indicate the highest-priority
active alarm condition.
7.5 Checking the alarm system
To test the alarm system:
1. In standby mode, disconnect the breathing tube
then press “Start”.
2. Verify that the “Check tube” visual alarm appears
on screen.
3. Verify that the signal light fl ashes yellow.
4. Verify that the auditory alarm signal can be
heard.
Do not use the Airvo 3 if it fails this test. Contact
your Fisher & Paykel Healthcare representative.
7.6 Airvo 3 alarms
All the alarms you may encounter while using the
Airvo 3 are listed below with common causes, resolutions and any delays inherent in determining alarm conditions. These priorities have
been allocated for an intended operator’s position of 2 meters from the device. The Airvo 3 also uses an internal-priority ranking system.
If multiple alarm conditions occur simultaneously, the device will display the highest-priority alarm.
Alarm condition
Priority
Delay
Meaning
Faults
Device Fault
[Fault X.X.X]
High
-
A technical fault has occurred that has caused therapy to be stopped. Restart
the device to resolve the fault. If the issue persists, contact your service
representative.
Device Fault
[Fault X.X.X]
Med.
-
A technical fault has occurred, and the device is able to continue supplying
limited therapy. Restart the device to resolve the fault. If the issue persists,
contact your service representative.
Power alarms
Power Out
High
<5 s
The Airvo 3 has been disconnected from the wall power supply and the internal
battery is depleted. The auditory alarm will sound once every 10 seconds for
120 seconds and the signal light above the touch screen will fl ash. The touch
screen is off during the Power Out alarm. The Airvo 3 will shut down after
signaling the Power Out alarm but will restart automatically if power is restored
before it shuts down.
Number of
active alarms
Active alarms
Suggestions/
action
Audio pause
Содержание Airvo 3
Страница 1: ...www fphcare com USER MANUAL...
Страница 2: ...SECTION English A Espa ol B C Fran ais Canadien D...
Страница 116: ...Airvo 3 C 7 3 Airvo 3 3 1 3 1 Airvo 3 AirSpiral Optiflow Airvo 3...
Страница 117: ...C 8 Airvo 3 3 2 HPO LPO USB 1 USB 2 HPO DISS NIST SIS...
Страница 118: ...Airvo 3 C 9 3 3 Airvo 3 3 3 1...
Страница 119: ...C 10 Airvo 3 3 3 2 7 3 3 3 Airvo 3 50 50 Airvo 3 USB USB 3 3 4 7...
Страница 123: ...C 14 Airvo 3 H2O 50 cm 5 Airvo 3 50 cm USP 6 MR290 7 Airvo 3 Fisher Paykel Healthcare Airvo 3 2...
Страница 126: ...Airvo 3 C 17 Airvo 3 1 2 1 Airvo 3 Airvo 3 2...
Страница 128: ...Airvo 3 C 19 1 3 5 3 Optiflow 1 3 2 1 2 3 Airvo 3 1 37 C 10 60 L min 31 C 1 2 3 7 5 1 Airvo 3 2 2 2 1...
Страница 130: ...Airvo 3 C 21 Airvo 3 EEG ECG EKG EMG Airvo 3...
Страница 131: ...C 22 Airvo 3 5 4 Airvo 3 10 120 Airvo 3 5 4 1 1 2 3 5 4 2 5 4 3 1 2 30 L min 1 2 3 3...
Страница 133: ...C 24 Airvo 3 5 6 1 2 Airvo 3 Airvo 3 1 2 3 2 Airvo 3 4 Airvo 3 Airvo 3 2 3 4 1 2...
Страница 134: ...Airvo 3 C 25 6 Airvo 3 24 Airvo 3 Airvo 3 Airvo 3 F P Airvo 3 F P Airvo 3 Airvo 3 Airvo 3...
Страница 136: ...Airvo 3 C 27 7 Airvo 3 Airvo 3 7 1 Airvo 3 Airvo 3 7 2 9 3 5 10 Airvo 3 7 3 Airvo 3 5 2 10 3 3...
Страница 141: ...C 32 Airvo 3 5 Airvo 3 35 Airvo 3 10 Airvo 3 1 Airvo 3 18 C 28 C Airvo 3 5 1 Airvo 3...
Страница 146: ...Airvo 3 C 37 9 9 1 USB Airvo 3 USB Airvo 3 9 1 1 Airvo 3 SpO2 9 2 USB Airvo 3 USB SpO2 Airvo 3 3 USB...
Страница 147: ...C 38 Airvo 3 SpO2 SpO2 Nonin Nonin Xpod USB Cardiogreen SpO Airvo 3 7 SpO2 SpO2...
Страница 149: ...C 40 Airvo 3 9 3 2 PR bpm 9 3 3 9 3 4 Nonin Nonin Airvo 3 SpO2 9 4 9 5 9 4 1 SpO2 SpO2 SpO2...