6. Troubleshooting
Follow the actions described in this chapter to inspect the instrument and troubleshoot the
problem if the load does not operate normally. Contact Faithtech authorized agent or Faithtech
after-sales if the information provided in this manual does not resolve the problem.
Table 6-1 Load fault check table
Problem
Possible Cause
Solution
Measurement accuracy is not
within SPEC.
Measurement swings due to
aged components.
Do calibration.
Loading accuracy is not within
SPEC.
Output swings due to aged
components.
Do calibration
Overtemperature Protection
OT
1. The ambient temperature
is too high.
2. Bad ventilation.
3. Fan damaged.
1. Operate the instrument within an ambient
temperature of 0 - 40°C.
2. Ensure the instrument is well ventilated.
3. Contact Faithtech authorized agent or
Faithtech after-sales.
Overpower Protection OPP
Load power exceeds set
Decrease loading or increase OPP setting
Overcurrent Protection OCP
Load current exceeds set
Decrease loading or increase OCP setting
Overvoltage Protection OVP
Load voltage exceeds set
Decrease input voltage or increase OVP
setting
Temperature Switch E01
1. The ambient temperature
is too high.
2. Bad ventilation.
3. Fan damaged.
1. Operate the instrument within an ambient
temperature of 0 - 40°C.
2. Ensure the instrument is well ventilated.
3. Contact Faithtech authorized agent or
Faithtech after-sales.
Fan error E02
Fan damaged.
1
.
Restart the load
2
.
Contact authorized dealer or Faithtech
after-sale service
Table 6-2 Start-up Fault
Problem
Possible Cause
Solution
Displays below message only:
(c)All rights reserved.
Internal communication error
1
.
Restart the load
2
.
Contact authorized dealer or
Faithtech after-sale service
Test sample.Error
!
Sampling data error
Loading calibration data. Error
!
Loading calibration data
failure
Loading parameters. Error
!
Loading system parameter
failure
Содержание FT68200 Series
Страница 9: ...Faithtech FT68200 Operation Manual 3 1 3 2 7U 8kW 10kW 12kW...
Страница 10: ...Faithtech FT68200 Operation Manual 4 1 3 3 11U 14kW 16kW 18kW...
Страница 11: ...Faithtech FT68200 Operation Manual 5 1 3 4 14U 20kW 22kW 24kW...
Страница 12: ...Faithtech FT68200 Operation Manual 6 1 3 5 20U 30kW 36kW...
Страница 13: ...Faithtech FT68200 Operation Manual 7 1 3 6 26U 42kW 48kW...
Страница 14: ...Faithtech FT68200 Operation Manual 8 1 3 7 32U 54kW 60kW...