Set Resistance Limit
FT68200 provides "resistance upper limit" and "resistance lower limit" setting items. When
these two parameters are set to non-zero values, the resistance setting range in constant
resistance mode is limited between the upper limit and the lower limit. This function can
prevent user misoperation and protect the device under test.
1. Press
(Menu) to go to main menu;
2. Select “Set”->“Limit” sub menu, press
to go to limit setting pape;
3. Select “R_MAX” and “R_MIN”, to set the desired value.
3.2.4. Constant Power Mode
Press
to switch to CP mode.
Figure 3-7 CP Mode Program
In CP mode, the load will sink a current according to the programmed power.
Figure 3-8 CP Mode
CP mode has only one range. “CP” will be prompted when load operates in CP mode.