Faithtech FT68200 Operation Manual
57
Operation Procedures
1. Press
(Test) to go to “Test” function selection;
2. Rotate the knob to select “DC_R”, press
and screen displays as follows:
Figure 3-24 Battery Internal Resistance Test
3. Configure the parameters;
4. Press
to start battery internal resistance test;
5. The “Rdc 12.5m
Ω
” display item in the readback area is the test result, i.e. the
battery internal resistance. PASS means that the internal resistance result is
between the upper limit and the lower limit of judgment, and FAIL means that
the test result exceeds the upper limit or the lower limit of judgment. If the
judgment upper limit or judgment lower limit is set to 0, it means that no
inspection will be performed, and PASS will be prompted directly.
Parameter Explanation
Name
Explain
CC-1
Test current 1
Width-1
Test current 1 continuous load time, the range is 1
~
60.0 seconds
CC-2
Test current 2
Width-2
Test current 2 continuous load time, the range is 1
~
60.0 seconds
Max Trip
The judgment upper limit of the internal resistance result, the setting
range is 0.000
~
99999m
Ω
Min Trip
The judgment lower limit of the internal resistance result, the setting
range is 0.000
~
99999m
Ω
Notice: If the remote sense cable or current cable is not in good contact, an
error will occur during the test and the screen will display ERROR.
Содержание FT68200 Series
Страница 9: ...Faithtech FT68200 Operation Manual 3 1 3 2 7U 8kW 10kW 12kW...
Страница 10: ...Faithtech FT68200 Operation Manual 4 1 3 3 11U 14kW 16kW 18kW...
Страница 11: ...Faithtech FT68200 Operation Manual 5 1 3 4 14U 20kW 22kW 24kW...
Страница 12: ...Faithtech FT68200 Operation Manual 6 1 3 5 20U 30kW 36kW...
Страница 13: ...Faithtech FT68200 Operation Manual 7 1 3 6 26U 42kW 48kW...
Страница 14: ...Faithtech FT68200 Operation Manual 8 1 3 7 32U 54kW 60kW...