Faithtech FT68200 Operation Manual
45
Operation Procedures
1. Press
(Menu) to go to main menu;
2. Select “Set”->“Protect”, press
to go to protect set;
3. Set “I_LIMIT”, press
to set current limit value.
3.12. Voltage remote sense
S+ and S- are remote sense terminals, used for electronic load sampling system to
provide remote voltage signal.
When it is necessary to accurately measure the output voltage of the DUT, it is
recommended to set the load as remote sense. The load terminals S+ and S- are directly
connected to the positive and negative output terminals of the DUT respectively to
eliminate the voltage drop on the wire, thereby obtaining a higher measurement accuracy.
Operation Procedures
1. Press
(Menu) to go to main menu;
2. Select “Set”->“APG”, press
to pop up application set page;
3. Set “Sense” to “remote” or “local”.
Notice:
If remote sense is selected for V-sense, but the load terminals S+ and S- are
not connected to the output terminal of the DUT, the electronic load will not be able to
measure the voltage correctly, and the constant voltage, constant resistance and constant
power modes will not work.
3.13. Current Monitoring
There is a
terminal, located on the rear of the device. It offers 0...10 V output
voltage to monitor load input current from 0...100%F.S. The accuracy of output voltage
Содержание FT68200 Series
Страница 9: ...Faithtech FT68200 Operation Manual 3 1 3 2 7U 8kW 10kW 12kW...
Страница 10: ...Faithtech FT68200 Operation Manual 4 1 3 3 11U 14kW 16kW 18kW...
Страница 11: ...Faithtech FT68200 Operation Manual 5 1 3 4 14U 20kW 22kW 24kW...
Страница 12: ...Faithtech FT68200 Operation Manual 6 1 3 5 20U 30kW 36kW...
Страница 13: ...Faithtech FT68200 Operation Manual 7 1 3 6 26U 42kW 48kW...
Страница 14: ...Faithtech FT68200 Operation Manual 8 1 3 7 32U 54kW 60kW...