6.0 Installation
Before connecting any components, follow the
safety information in the ETS-Lindgren
Product Information Bulletin included with your
shipment.
Prior to performing any of the steps in this
chapter, see Before You Begin Installation on
page 27.
The OMS must be installed outside of the
MRI room; otherwise, the OMS may interfere
with the normal operation of the MRI system.
The OMS must be installed outside of the shielded enclosure.
Mounting the OMS
The Sample Draw Oxygen Monitoring System (OMS™) is designed primarily for
wall mounting and should be installed at a height convenient for operation,
maintenance, and viewing of the instrument display. The transmitter and sensor
should be installed in a location where gas leaks are likely to occur or where
released gases may accumulate.