Required Gas and Equipment
The required gas and equipment are available from your gas supplier
or from Air Liquide at 800-638-1197.
Nitrogen and an on-demand regulator are recommended to expose the OMS to a
known span gas. You may purchase both items from your gas supplier or from
Air Liquide.
Air Liquide
Part Number
Description
CZF6D400281
Nitrogen 105 liter cylinder, 99.99%
18102509
Regulator, On-Demand
Sensor Verification Procedure
Prior to performing the procedure, place the OMS in a clean, non-oxygen
deficient environment.
Warm up the OMS for two hours prior to making any adjustments to
the sensor.
If the OMS is connected to a controller, put the controller in standby
mode to avoid accidental alarms.
If required by your safety protocol, you may subject the OMS to
different concentrations of oxygen span gas.