27
www.ets-lindgren.com
Before You Begin Installation
5.0 Before You Begin Installation
Before installing or connecting any
components, follow the safety information in
the ETS-Lindgren Product Information Bulletin
included with your shipment.
Failure to follow the instructions in this chapter
may damage the OMS and/or the sensor.
Site Requirements
The OMS is not designed for installation in a
hazardous area. Contact ETS-Lindgren for
information on special enclosures for use in a
hazardous environment.
The OMS must be installed outside of the
MRI room; otherwise, the OMS may interfere
with the normal operation of the MRI system.
The OMS must be installed outside of the shielded enclosure.
The Sample Draw Oxygen Monitoring System (OMS™) enclosure should be
mounted in an area free of vibration and electrical noise or interference. If
possible, avoid areas with high temperatures or condensing humidity.
Содержание OMS