Programming the OMS
www.fmgrupotec.com
0034.903.360.306 Calle Algepser n16 Parque Empresarial Tactica- 46980 Paterna (Valencia)- España
S
ET
A
LARM
D
ELAY
The Alarm Delay is the amount of time an alarm level concentration of oxygen
must be present before the gas concentration alarm(s) will be activated.
Use this menu to set a user-selected time delay for activating
alarm relays 1 and 2, from 0 seconds to 255 seconds. When an alarm level has
been exceeded, the alarm relays will activate after this period of time.
Step
Display
1. Push the joystick down to
access the Alarm Delay
main menu.
Alarm Delay
2. Push the joystick to the right to
display the time setting
function. Push the joystick up
to increase the value or down
to decrease the value, and
then Enter to accept the value.
The display will indicate a value
between 0 and 255 seconds, and
then will default to the Alarm Delay
menu.
The alarm delay is only available for alarm 1 and alarm 2; there is no
delay for the fault relay. Any system fault will immediately activate the
fault relay.
S
ET
Z
ERO
S
UPPRESSION
Although these settings can be changed, this function is disabled on the OMS; it
is only used to decrease the sensitivity of selected gas sensors. The factory
default is set at 000.
Содержание OMS
Страница 8: ......
Страница 10: ...This page intentionally left blank ...
Страница 14: ...This page intentionally left blank ...
Страница 18: ...16 Maintenance This page intentionally left blank ...
Страница 26: ...24 Component Views Front view with electronics panel rotated ...
Страница 28: ...26 Component Views Alarm Relay Board For the relays from left to right NC C NO ...
Страница 43: ...41 www ets lindgren com Installation ...
Страница 48: ...46 Installation www ets lindgren com This page intentionally left blank ...