Sensor Verification
www.ets-lindgren.com
www.fmgrupotec.com
0034.903.360.306 Calle Algepser n16 Parque Empresarial Tactica- 46980 Paterna (Valencia)- España
Sensor Verification to Known Concentration of Oxygen
Protect the OMS from wind and high airflow
when gas calibrating with test gas.
Test the OMS in an upright position to allow the span gas to fully
saturate the sensor.
To see the exact span gas concentration the OMS must be
completely immersed into a span gas environment.
To see a true zero, the OMS must be completely immersed into a
zero oxygen environment.
The sensor inlet on the OMS has a 1/4-in compression tube fitting designed to
connect to the filter. Connect 1/4-in diameter sample tubing from a nitrogen
cylinder directly to the filter. Expose the OMS directly from the nitrogen cylinder
and the reading will drop off to the span gas concentration in less than
one minute. The final reading should be within ±0.3% of the span gas
concentration.