Programming the OMS
www.ets-lindgren.com
www.fmgrupotec.com
0034.903.360.306 Calle Algepser n16 Parque Empresarial Tactica- 46980 Paterna (Valencia)- España
S
ET
S
ENSOR
A
DJUST
Warm up the OMS for two hours prior to making any adjustments to
the sensor.
Sensor Adjust sets the oxygen readout to a known concentration. It is
recommended to adjust the oxygen display to ambient oxygen levels of 20.9%.
Step
Display
1. Push the joystick down to
access the menu.
Sensor Adjustment
2. Push the joystick to the right to
access the submenu.
Set Sensor Span
3. Push the joystick up to
increase the counts and
decrease the percent oxygen
value displayed. Push the
joystick down to decrease the
counts and increase the
oxygen value displayed. Adjust
to a value of 20.9% ± 0.1%,
and then Enter to accept the
value.
The display will indicate a value
between 0 and 255 counts, and then
will display the selected value of
20.9% ± 0.1%. Then the display will
default to Set Sensor Span.
The actual oxygen reading will fluctuate from 20.8% to 21.0%
4. Push the joystick to the left to
go back to the main menu.
Sensor Adjustment
Set Module Zero is not available for the OMS; it was designed for
toxic and corrosive gas monitors. Push the joystick to the left to exit
this menu.
Содержание OMS
Страница 8: ......
Страница 10: ...This page intentionally left blank ...
Страница 14: ...This page intentionally left blank ...
Страница 18: ...16 Maintenance This page intentionally left blank ...
Страница 26: ...24 Component Views Front view with electronics panel rotated ...
Страница 28: ...26 Component Views Alarm Relay Board For the relays from left to right NC C NO ...
Страница 43: ...41 www ets lindgren com Installation ...
Страница 48: ...46 Installation www ets lindgren com This page intentionally left blank ...