Operation
Instrument Faults
When using your own power supply, make sure that the voltage is
regulated to 24 VDC ± 0.5 Volts. A voltage that is too low or too high
will activate a Supply Voltage Out of Range fault and disable
the OMS.
The OMS incorporates a number of self-checking features to ensure reliable
operation. If a fault condition is detected, the analog output signal is altered as
follows:
All system faults are displayed on the front panel. Each fault has a
unique code to identify the specific problem. Contact ETS-Lindgren if
a fault is displayed.
Condition
Analog Signal
Supply Voltage Out of Range
Fault Code 16
Analog output drops to 2 mA
Transmitter cable cut
Analog output drops to 0 mA
O
2
Cell Current Fault
Fault Code 128
Analog output drops to 2 mA
(0 mA on request)
Fault Relay activates
(available with Relay Option only)
O
2
System Warm Up
Analog output drops to 2 mA
Fault Relay activates and turns off
when system is in Oxygen
Operation Mode
O
2
Cell Voltage Fault
Fault Code 64
Analog output drops to 2 mA
Fault Relay activates
(available with Relay Option only)
No Flow to Oxygen Sensor
Fault Code 32
Analog output drops to 2 mA
Fault Relay activates
EEPROM Fault 08
Analog output drops to 2 A
(0 mA on request)
Содержание OMS
Страница 8: ......
Страница 10: ...This page intentionally left blank ...
Страница 14: ...This page intentionally left blank ...
Страница 18: ...16 Maintenance This page intentionally left blank ...
Страница 26: ...24 Component Views Front view with electronics panel rotated ...
Страница 28: ...26 Component Views Alarm Relay Board For the relays from left to right NC C NO ...
Страница 43: ...41 www ets lindgren com Installation ...
Страница 48: ...46 Installation www ets lindgren com This page intentionally left blank ...