63
www.ets-lindgren.com
Programming the OMS
www.fmgrupotec.com
0034.903.360.306 Calle Algepser n16 Parque Empresarial Tactica- 46980 Paterna (Valencia)- España
S
ET
A
LARM
T
HRESHOLD
P
ARITY
Alarm Threshold Polarity determines if an alarm concentration is set above or
below a threshold value. For example, if an alarm of 19.0% for oxygen is
selected, the Alarm Threshold Polarity must be set to INVERT for the alarm to
activate when the reading goes below 19.0%.
For toxic and corrosive gases, a NORMAL setting for the Alarm Threshold
Polarity causes the system to alarm when the gas concentration
exceeds/goes above an alarm set point.
Step
Display
1. Access this menu from the
Set Formats menu by pushing
the joystick down.
Set Alarm Threshold Polarity
2. Push the joystick to the right.
Set Alarm 1 Polarity
3. Push the joystick to the right to
display the relay state.
INVERT
4. Push the joystick down to
change the relay state to
NORMAL, and then Enter to
accept the value.
NORMAL will display, and then will
default to the Set Alarm 1 Polarity
menu.
5. Push the joystick down to
access the next submenu.
Set Alarm 2 Polarity
6. Push the joystick to the right to
display the relay state.
INVERT
7. Push the joystick down to
change the relay state to
NORMAL, and then Enter to
accept the value.
NORMAL will display, and then will
default to the Set Alarm 2 Polarity
menu.
Содержание OMS
Страница 8: ......
Страница 10: ...This page intentionally left blank ...
Страница 14: ...This page intentionally left blank ...
Страница 18: ...16 Maintenance This page intentionally left blank ...
Страница 26: ...24 Component Views Front view with electronics panel rotated ...
Страница 28: ...26 Component Views Alarm Relay Board For the relays from left to right NC C NO ...
Страница 43: ...41 www ets lindgren com Installation ...
Страница 48: ...46 Installation www ets lindgren com This page intentionally left blank ...