Programming the OMS
www.fmgrupotec.com
0034.903.360.306 Calle Algepser n16 Parque Empresarial Tactica- 46980 Paterna (Valencia)- España
Step
Display
6. To change the value, push the
joystick to the right to display
the 20 mA span setting.
The display will indicate a value
between 0 and 255 counts.
7. Push the joystick up to
increase the value and push it
down to decrease the value;
Enter to accept the value.
The 20 mA output sent from
the OMS will change as the number
on the display changes.
8. Push the joystick to the left to
go back to the previous
main menu.
Set 20mA Span
You will be able to access the Force Loop submenu, but because
the Force Loop function was designed for toxic and corrosive gases,
it is not available for use on the OMS.
S
ET
F
ORMATS
This set of menus is used to adjust the relay states for the two gas alarm relays
and the individual instrument fault relay.
To access this menu, the OMS must be installed with the relay
module. If not installed, the display will indicate N/A (not available).
Step
Display
1. Push the joystick down to
access the next main menu,
Set Formats. This is used to
set the two alarm relays and
the fault relay settings from a
normally de-energized state
(Normal) to a normally
energized state (Inverted).
Set Formats
Содержание OMS
Страница 8: ......
Страница 10: ...This page intentionally left blank ...
Страница 14: ...This page intentionally left blank ...
Страница 18: ...16 Maintenance This page intentionally left blank ...
Страница 26: ...24 Component Views Front view with electronics panel rotated ...
Страница 28: ...26 Component Views Alarm Relay Board For the relays from left to right NC C NO ...
Страница 43: ...41 www ets lindgren com Installation ...
Страница 48: ...46 Installation www ets lindgren com This page intentionally left blank ...