NOTE:
Calibration can ONLY be performed by a User at the Administrator Permission level.
Place the Calibration Tray into one of the tray positions (the order in which tray positions are
calibrated is not important, but all 3 must be calibrated).
Navigate to the Tests tab under the Administer IncuCyte™ task bar. The Front, Middle and Rear
tray positions can be selected using the pull-down menu. Each Tray Position must be individually
calibrated and confirmed. The first time the instrument is calibrated, select the “Confirm Test”
check box to run the Calibrate and Confirm Test (refer to Figure 12).
The status if the Calibration and Confirm test completed successfully will be displayed in the far
right column of the table labeled “Result” (pass/fail status is reported for each magnification and
tray position combination). You might need to click on the green “Refresh” button on the bottom
to see the updated status. A summary of calibration details and accompanying images can be
accessed via the “View Test Results” button located at the bottom of the screen. The calibration
results will be displayed in a new window. The results will be displayed as coordinate values. The
Log will also indicate whether the test passed or failed. The detailed results of the Calibration
and Confirm test can also be viewed in the Logs tab. Select Logs, and then use the pull-down
Figure 12.
Calibration