Figure 7.
Microscope with filter module installed
3.4 Installing an Objective
The IncuCyte™ ZOOM ships with up to three objectives with magnifications of 4X, 10X and 20X.
The system will need to be calibrated for all of the purchased objectives. To get started, the 10X
objective should be installed if it is available, otherwise the 20X or 4X should be used. Objectives
are installed on a bracket that is shipped already installed in the system. See the diagram above.
With the filter module access cover removed, pull out the drawer halfway to provide access.
Remove the objective bracket from the system by loosening the screw using the supplied 2.5mm
T handle hex wrench and then pulling the bracket straight out. Finger tighten the objective into
the bracket. Install the bracket and objective back into the system on the two guide pins and
retighten the screw using the 2.5 mm T handle hex wrench. If you are installing a 20X objective,
verify that the corrective ring on the objective is set to 0.8. Close the drawer and replace the
filter module access cover.
3.5 Positioning the Controller
It is important to put the controller into position before placing the microscope into the
incubator. The controller should be placed on a level surface within 3 feet of the incubator's