Page 24
IMPORTANT:
Always use the “Disconnect” option and not the “Log Off” option when
terminating the Remote Desktop Connection via the controller Start menu.
Run the “Remote Desktop Connection” software and type in the name of the IncuCyte™ ZOOM
controller in the “Computer” edit box. If connecting for the very first time, the name used will be
the factory-set name. Make sure that you have connected and verified that the controller is on
the network (see Sections: 4.1.1 and 4.1.2). Press the “Connect” button and wait for a Windows
login prompt. It may take as long as 15 seconds before the prompt appears. Log into the
controller with a User name "ZoomAdmin" and the Essen-provided password. The controller
operating system should contain the necessary components required to rename the computer,
configure the primary Ethernet port, install security software, etc. Please consult with your IT
department for answers to questions related to your local network configuration and security
software.
Chapter 6: Calibrating the IncuCyte
™
ZOOM System
6.1 System Calibration
Before using the IncuCyte™ ZOOM for the first time, a calibration procedure must be performed.
This procedure ensures that the motion of the IncuCyte™ ZOOM microscope is accurate and
repeatable such that images are obtained within their proper, target locations. This is especially
important when acquiring images in 96 and 384-well plates. The calibration procedure requires
up to 1.5 hours, and it utilizes a special calibration tray that is shipped out with the unit (Figure
11). The calibration tray is the same size and is inserted in the same fashion as the regular trays.
Each of the three tray positions (front, middle and rear) must be calibrated individually. What
follows is an overview of the calibration procedure.
Figure 11.
The Calibration Tray