Criterion Series Benchtop User’s Manual
43
7.3. User Practical Servicing
7.3.1. Resetting Circuit Protector
When a circuit protector trips it can be reset manually.
Note
If a circuit protector trips soon after being reset, contact the place of purchase or ESPEC.
Procedure
1. Unplug Chamber from power source.
!
DANGER
•
Turning controller power off does not make the chamber safe to open the electrical box
or breaker access area. The only safe method of access is to first unplug the chamber
from the power receptacle.
2. Open the circuit protector access panel (Phillips screwdriver required) Locate removable
panel per section 3.1.3.
View of circuit protectors in the “ON” position
3. Located the tripped circuit protector. To reset move the protector to the off position and
move back to the on position.
4. Secure access cover with all fasteners.