Criterion Series Benchtop User’s Manual
26
4.4.1. Water Supply Installation
Connection of the Water Supply Hose for the Humidifying Water Circuit
!
CAUTION
•
Remove packing sponges from the float level cups before supplying water to the
chamber. Failure to do so can cause water to overflow in the humidity system and
spill water on your laboratory floor making the floor extremely slippery.
Attach supply water to the water supply fitting on the left side of the chamber. The supply water
must be from a purified* water source. The water supply should have a pressure between 20 ~
50 psi to assure proper operation of the humidifying circuit.
*The type of purifying system does not matter, but the purified water should be filtered (5
micron), have a conductivity of 0.2µS/cm to 10µS/cm, and a maximum of 2mg/L of free chlorine.
If your chamber is equipped with an optional water supply tank please see the option manual for
the water supply installation procedure.
4.4.2. Water Consumption Rate
CHAMBER MODEL
WATER CONSUMPTION
RATE
BTL-433 , BTX-475
Peak* 5 GPH/Typical 0.3 GPH
BTL-133 , BTX-133
Peak* 3 GPH/Typical 0.2 GPH
*Peak flow rates are for the initial filling of the humidifier or for filling after a temperature only set point.
4.4.3. Water Level Setting
4.4.3.1. Checking Water Level in the Wick Pan
Procedure
1. The standard water level in the wick pan arm should be set so that water
does not overflow to the chamber, or, is not extremely low.
2. The wick pan arm should be level.
Fig 4.4.3.
Box Nut
Valve Seat
Valve
Nut
Содержание BTL-433
Страница 1: ...CRITERION CHAMBER SERIES BENCH TOP INSTRUCTION MANUAL...
Страница 17: ...Criterion Series Benchtop User s Manual 14...
Страница 23: ...Criterion Series Benchtop User s Manual 20...
Страница 31: ...Criterion Series Benchtop User s Manual 28...
Страница 41: ...Criterion Series Benchtop User s Manual 38...
Страница 48: ...Criterion Series Benchtop User s Manual 45 8 Appendix 8 1 Specifications...
Страница 56: ...Criterion Series Benchtop User s Manual 49...