Criterion Series Benchtop User’s Manual
22
4.2. Power Supply Work
4.2.1. Primary Power Supply Requirements
The primary power supply must satisfy the following requirements.
•
Voltage fluctuation must be within ± 5% of the rated voltage. (Within ± 10% while the
chamber is running.).
•
A power cable is provided of the proper wire gauge for the chamber and an appropriate
plug is also provided. The following table gives the plug type provided with the chamber so
you may provide the proper receptacle.
MODEL Plug
type
BTL-133 NEMA
5-20
BTL-433 NEMA
5-20
BTX-133 NEMA
5-20
BTX-475 NEMA
6-15
BTZ-133 NEMA
5-20
BTZ-175 NEMA
6-15
BTZ-475 NEMA
6-15