Criterion Series Benchtop User’s Manual
42
Condition Cause Remedial
Action
The door is not “latched”.
Shut the door. See section 5.1.2.
Ambient temperature changes
more than 5°C/hr.
Stabilize ambient temperature and
resume testing.
Temperature (& Humidity) is
unstable.
High heat load equipment is being
turned ON/OFF.
Reduce the heat load.
Temperature gradually rises
higher than the target
temperature.
Specimens are generating heat.
Reduce the amount of heat
generated by specimens.
The door is not “latched”.
Shut the door. See section 5.1.2.
Specimens are generating heat.
Reduce the amount of heat
generated by specimens.
Ambient temperature too low
(high).
Raise (lower) ambient temperature.
Temperature rises (lowers)
too slowly.
Circuit protector for compressor or
heater is tripped.
Reset circuit protector per section
7.3.1. If it trips a second time call
ESPEC for service
Air flow inside the chamber is
poor.
Improve air flow by rearranging
specimens. Be sure not to block the
air supply or return. See section
3.1.4. for location of air supply and
return.
Poor temperature uniformity
Specimens are generating heat.
Reduce the amount of heat
generated by specimens.
The chamber lamp does not
light up.
Light bulb is burned out
Replace bulb. See “7.3.2
Содержание BTL-433
Страница 1: ...CRITERION CHAMBER SERIES BENCH TOP INSTRUCTION MANUAL...
Страница 17: ...Criterion Series Benchtop User s Manual 14...
Страница 23: ...Criterion Series Benchtop User s Manual 20...
Страница 31: ...Criterion Series Benchtop User s Manual 28...
Страница 41: ...Criterion Series Benchtop User s Manual 38...
Страница 48: ...Criterion Series Benchtop User s Manual 45 8 Appendix 8 1 Specifications...
Страница 56: ...Criterion Series Benchtop User s Manual 49...