4.3. Drainage
!
CAUTION
•
If collecting drainage in a container, check water level to make sure it does not
overflow.
Drainage is needed to drain water from the humidifier (BTX and BTL models only) and dew formed in
the test area out of the chamber.
Procedure
Connect a hose fitting and hose to the threaded drain fitting.
The drain fitting size is 3/8” FPT.
The hose used for drainage must have a minimum internal diameter of 3/8”.
Note
If a drainage pit is not located near enough to the chamber, run drainage into a container or
condensate pump. Whether running drainage into a pit or a container, keep the end of the drain
hose open to the atmosphere. If submerged, water may backflow into the chamber.