Criterion Series Benchtop User’s Manual
40
Clearing Alarms
Please refer to the Watlow Manual for instructions.
7.2 Other Troubles
DANGER
•
ELECTRIC SHOCK! Before working on the power
circuits, unplug chamber. Also, take measures to
prevent accidental charging.
Working with primary power supply ON runs the risk of electric
shock.
•
Unplug power from the chamber BEFORE opening
the electrical compartment panel.
This section explains troubles undetected in cabinet self-checks and cases of misoperation which are easily
mistaken as trouble. If the trouble cannot be remedied after taking the prescribed action, contact the place of
purchase or ESPEC.
Table 7.2 Other troubles
Trouble Cause Remedial
action
The instrumentation
panel does not light up
after pressing the
POWER key.
Primary power supply is OFF.
Activate the primary power supply.
The display goes out all
of sudden or the
displayed information is
strange.
System trouble or internal board
trouble
Remove power from the chamber by
unplugging from receptacle. Reapply
power. If the same trouble reoccurs,
call for service.