27
2. Attach the extension to the tension scale,
and then insert the end of the extension into the
tensioner aperture located on the belt tensioner.
3. Pull the belt tensioner to the right with the scale until the scale reads 48 oz. This is the maximum belt tension
recommended.
4. Holding the scale steady at 48 oz, re-tighten
the belt tensioner bolts.
When all tensioner bolts are tightened, remove the extension from the aperture and place the scale in a safe,
retrievable location.