1.1.2.2
Electrical and pneumatic connections
1.1.2.2.1
Fixed tight box version (Figure 1–3)
The gas inlets/outlets
–
The "SAMPLE OUTLET" for connection to an open air venting system, if necessary, is made up
with a 4/6 mm PVDF connection (1).
–
The "SAMPLE" inlet is made up with a 2/4 mm Teflon connection or with a 4/6 mm PVDF connec-
tion (2).
–
The "ZERO AIR" inlet for connection, at less than 10
3
hPa pressure, to a nitrogen gas cylinder or to
a dry and oil-free air instrument, is made up with a 4/6 mm PVDF connection (3). It can be also
connected to the automatic calibration module (TIG).
–
The "SPAN" inlet, for connection to a gas cylinder or to a manual or automatic calibration module
(TIG), is made up with a 4/6 mm PVDF connection (4).
–
The outlet (5) of the oxygen cell (optional) is made up with 4/6 mm PVDF connections.
Electrical connections and equipment
–
A
13 mm packing box (7) enables the passing of a 26-conductor cable for the connection to TIG.
–
A
16 mm packing box (8) enables the passing of a 60-conductor cable from the analog, digital
and logic outputs (towards the interface board).
–
A
13 mm packing box (6) enables the passing of the mains power supply cable,
–
A RJ45 socket (9) used for Ethernet connection.
1.1.2.2.2 NH
3
/H
2
O option - tight box version (Figure 1–4)
The gas inlets/outlets
–
The gas connection between the MIR 9000 tigh box and the sample is made through the heated
line (2). There is only one gaseous inlet for the "SAMPLE", "ZERO AIR" or "SPAN". Gases connec-
tions and solenoid valves are installed upper in the tight dilution box. The configuration allows a
span check of the complete measuring system.
–
The "SAMPLE" inlet is fitted with an antidust filter (7 µm) integrated under the measurement cham-
ber upper heating plate.
–
The "SAMPLE OUTLET" for connection to an open air venting system is made up with a 4/6 mm
PVDF connection (1).
–
The air inlet for the ejector is made up with a 4/6 mm PVDF(3).
Electrical connections and equipment
–
A
13 mm packing box (9) enables the passing of the 26-conductor cable for the connection to
the TIG.
–
A
16 mm packing box (8) enables the passing of the 60-conductor cable from the analog, digital
and logic outputs (towards the interface board).
–
A
13 mm packing box (7) enables the passing of the mains power supply cable.
–
A RJ45 socket used for Ethernet connection (not represented on the picture).
Содержание Envea MIR 9000
Страница 10: ...Duplication prohibited MIR 9000 EnvironnementS A 0 10 JULY 2009 Page intentionally left blank...
Страница 22: ...EnvironnementS A MIR 9000 Duplication prohibited 1 12 JULY 2009 Figure 1 7 Analysis cabinet...
Страница 24: ...EnvironnementS A MIR 9000 Duplication prohibited 1 14 JULY 2009 Figure 1 8 Example of application...
Страница 27: ...Duplication prohibited MIR 9000 EnvironnementS A 1 17 JULY 2009 Page intentionally left blank...
Страница 35: ...Duplication prohibited MIR 9000 EnvironnementS A 2 5 JULY 2009 Page intentionally left blank...
Страница 39: ...Duplication prohibited MIR 9000 EnvironnementS A 2 9 JULY 2009 Figure 2 5 Basic fluid diagram of MIR9000...
Страница 44: ...EnvironnementS A MIR 9000 Duplication prohibited 2 14 NOVEMBER 2014 Page intentionally left blank...
Страница 102: ...EnvironnementS A MIR 9000 Duplication prohibited 4 2 JULY 2009 Page intentionality left blank...
Страница 146: ...EnvironnementS A MIR 9000 Duplication prohibited 5 28 JULY 2009 Figure 5 18 Motherboard configuration...
Страница 148: ...EnvironnementS A MIR 9000 Duplication prohibited 5 30 JULY 2009 Figure 5 19 Test points and settings of the motherboard...
Страница 150: ...EnvironnementS A MIR 9000 Duplication prohibited 5 32 JULY 2009 Figure 5 20 Configuration and settings of the DAC8 board...
Страница 157: ......
Страница 159: ......
Страница 161: ......
Страница 163: ...EnvironnementS A MIR 9000 Duplication prohibited 6 10 JULY 2009 Figure 6 4 Oxygen probe...
Страница 165: ...EnvironnementS A MIR 9000 Duplication prohibited 6 12 JULY 2009 Page intentionally left blank...
Страница 173: ...EnvironnementS A DNP ARM7 board Duplication prohibited APRIL 2008 8 Page intentionally left blank...
Страница 180: ...EnvironnementS A MIR 9000 Duplication prohibited 6 22 JULY 2009 Page intentionally left blank...
Страница 182: ...EnvironnementS A MIR 9000 Duplication prohibited 7 2 JULY 2009 Page left intentionally blank...
Страница 206: ...EnvironnementS A MIR 9000 Duplication prohibited 7 28 JULY 2009 Page left intentionally blank...
Страница 212: ...EnvironnementS A DAC8 board Duplication prohibited MARCH 2018 6 Page intentionally left blank...