Duplication prohibited
MIR 9000
Environnement
S.A
5–13
JULY 2009
5.2.5
ADJUSTMENT OF THE MOTOR SPEED
This should be adjusted by using potentiometer P12 and setting an identical value to that indicated on
the control board accompanying each analyzer.
In the “Tests / Optical Bench” menu, check the RPM value corresponding to the motor speed signal.
The motor speed signal can also be visualized in the “Measurement/Synoptic” menu.
5.2.6
ADJUSTMENT OF THE AMPLIFICATION CHAIN
To visualize the different signals, a double-trace oscilloscope is required. The grounds of the probes
should be connected to point Pt 25 of the motherboard.
5.2.6.1
Control of the optical forks.
−
“Signal chopper” optical fork.
Connect the probe of the oscilloscope to Pt 9 in order to visualize the signal. Its amplitude should
be > 3.800 mV. Otherwise, change the optical fork.
−
“Index GFC” optical fork
Connect the probe of the oscilloscope to Pt 13 in order to visualize the signal. Its amplitude should
be > 3.800 mV. Otherwise, change the optical fork.
After replacing the optical fork, it is essential to start up the control procedure again in order to make
sure of its characteristics.
Figure 5–1 – Oscillogram signals optical forks
Содержание Envea MIR 9000
Страница 10: ...Duplication prohibited MIR 9000 EnvironnementS A 0 10 JULY 2009 Page intentionally left blank...
Страница 22: ...EnvironnementS A MIR 9000 Duplication prohibited 1 12 JULY 2009 Figure 1 7 Analysis cabinet...
Страница 24: ...EnvironnementS A MIR 9000 Duplication prohibited 1 14 JULY 2009 Figure 1 8 Example of application...
Страница 27: ...Duplication prohibited MIR 9000 EnvironnementS A 1 17 JULY 2009 Page intentionally left blank...
Страница 35: ...Duplication prohibited MIR 9000 EnvironnementS A 2 5 JULY 2009 Page intentionally left blank...
Страница 39: ...Duplication prohibited MIR 9000 EnvironnementS A 2 9 JULY 2009 Figure 2 5 Basic fluid diagram of MIR9000...
Страница 44: ...EnvironnementS A MIR 9000 Duplication prohibited 2 14 NOVEMBER 2014 Page intentionally left blank...
Страница 102: ...EnvironnementS A MIR 9000 Duplication prohibited 4 2 JULY 2009 Page intentionality left blank...
Страница 146: ...EnvironnementS A MIR 9000 Duplication prohibited 5 28 JULY 2009 Figure 5 18 Motherboard configuration...
Страница 148: ...EnvironnementS A MIR 9000 Duplication prohibited 5 30 JULY 2009 Figure 5 19 Test points and settings of the motherboard...
Страница 150: ...EnvironnementS A MIR 9000 Duplication prohibited 5 32 JULY 2009 Figure 5 20 Configuration and settings of the DAC8 board...
Страница 157: ......
Страница 159: ......
Страница 161: ......
Страница 163: ...EnvironnementS A MIR 9000 Duplication prohibited 6 10 JULY 2009 Figure 6 4 Oxygen probe...
Страница 165: ...EnvironnementS A MIR 9000 Duplication prohibited 6 12 JULY 2009 Page intentionally left blank...
Страница 173: ...EnvironnementS A DNP ARM7 board Duplication prohibited APRIL 2008 8 Page intentionally left blank...
Страница 180: ...EnvironnementS A MIR 9000 Duplication prohibited 6 22 JULY 2009 Page intentionally left blank...
Страница 182: ...EnvironnementS A MIR 9000 Duplication prohibited 7 2 JULY 2009 Page left intentionally blank...
Страница 206: ...EnvironnementS A MIR 9000 Duplication prohibited 7 28 JULY 2009 Page left intentionally blank...
Страница 212: ...EnvironnementS A DAC8 board Duplication prohibited MARCH 2018 6 Page intentionally left blank...