Manual linearization with
tabular values
Enter you linearization values in the Table below: Channel 1 = V2H2
…
V2H4,
channel 2 = V6H2
…
V6H4
No.
V2H2
V6H2
Volume
V2H3
V6H3
Level
V2H4
V6H4
No.
V2H2
V6H2
Volume
V2H3
V6H3
Level
V2H4
V6H4
1
16
2
17
3
18
4
19
5
20
6
21
7
22
8
23
9
24
10
25
11
26
12
27
13
28
14
29
15
30
Procedure
Step
Matrix
Entry
Significance
1
V9H5/V3H5/V3H6
Reset and commission transmitter, see Section 4.1
2
V0H1/V0H2
…
Calibrate transmitter, see Sections 4.2
…
4.3 or Chapter 8
3
V2H0
4
Cancel any previously active linearization table
4
-
»E«
Register entry
5
V2H1
0
Select manual entry of values
6
-
»E«
Register entry
7
V2H2
1
…
30
Enter table entry number
8
-
»E«
Register entry
9
V2H3
e.g. 0
Enter volume in required units
10
-
»E«
Register entry
11
V2H4
00.00
Enter level in metres or feet
12
-
»E«
Register entry
13
V2H5
2
…
30
Enter next table entry number
14
-
»E«
Register entry.
-
The system jumps to V2H3; V2H2 is incremented
15
V2H3
Repeat steps 9 to 14 until all points are entered
16
V2H0
3
Select manual linearization table
17
-
»E«
Press »E« to activate linearization
Note!
•
Omit step 1, transmitter reset, when the second linearization is being entered
•
For channel 2 use fields V4H1/V4H2 and V6H0
…
V6H5.
Silometer FMC 672 Z/677 Z
Chapter 5: Linearization
30