Technical data
Dosimass
36
Hauser
10.1.7
Installation
Installation instructions
Inlet and outlet run
There are no installation requirements regarding inlet and outlet runs.
System pressure
10.1.8
Environment
Ambient temperature range
–20 to +60 °C (–4 to +140 °F) (sensor, transmitter)
Install the device at a shady location. Avoid direct sunlight, particularly in warm climatic regions.
Storage temperature
–40 to +80 °C (–40 to +175 °F), preferably at +20 °C (+68 °F)
Degree of protection
Standard: IP 67 (NEMA 4X) for transmitter and sensor
Shock resistance
In accordance with IEC 68-2-31
Vibration resistance
In accordance with IEC 68-2-31
CIP/SIP cleaning
Yes
Electromagnetic compatibility
In accordance with IEC/EN 61326
10.1.9
Process
Medium temperature range
Sensor:
• –40 to +125 °C (–40 to +257 °F)
CIP/SIP cleaning (< 60 min):
• +150 °C (+302 °F)
Seals:
• No internal seals
Medium density:
• 0 to 5000 kg/m
3
(0 to 312 lb/cf)
Medium pressure range
Max. 100 bar (1450 psi), depending on process connection
Limiting flow
See information on
33 ("Measuring range")
Select nominal diameter by optimizing between required flow range and permissible pressure loss.
See
33, "Measuring range" for a list of maximum possible full scale values.
• The minimum recommended full scale value is approx. 1/20 of the maximum full scale value.
• In most applications, 20 to 50% of the maximum full scale value can be considered ideal.
• Select a lower full scale value for abrasive substances such as fluids with entrained solids (flow
velocity < 1 m/s (3 ft/s)).