Appendix - Function description
Dosimass
44
Hauser
11.3
Function group PULSE OUTPUT
Function description
Function group PULSE OUTPUT
ASSIGN PULSE
Use this function to assign a measured variable to the pulse output.
Options:
OFF
MASS FLOW
VOLUME FLOW
Factory setting:
MASS FLOW
PULSE VALUE
Use this function to define the flow at which a pulse is triggered. These pulses can be
totaled by an external totalizer, and the total flow quantity since measuring started can
be registered in this way.
User input:
Multi-digit floating-point number
Factory setting:
SI units:
DN 8
0.02 g
DN 15
0.10 g
DN 25
0.20 g
US units:
DN 8
0.001 oz
DN 15
0.004 oz
DN 25
0.010 oz
!
Note!
The appropriate unit is taken from the SYSTEM UNITS function group (
42 ff.).