Troubleshooting
Dosimass
30
Hauser
9.9
Installing/removing the electronics
#
Warning!
Risk of damaging electronic components (ESD protection). Static electricity can damage Electronics
components or impair their operability. Use a workplace with a grounded working surface purposely
built for electrostatically sensitive devices!
1.
Switch off power supply.
2.
Loosen hexagonal-headed bolt (1) and remove electronics cover (2).
3.
Disconnect cable connectors from the electronics boards:
– Plug of the sensor signal cable (3) incl. DAT (4)
– Plug of the exciting current cable (5)
– Plug for power supply and signal outputs (6)
4.
Loosen Phillips screws (7) and remove Electronics module (8).
5.
Installation is the reverse of the removal procedure.
"
Caution!
Use only original Hauser parts.
A0004756
Fig. 17:
Installing and removing the electronics
1
Hexagonal-headed bolt (AF 10)
2
Electronics compartment cover
3
Plug of the sensor signal cable
4
DAT (sensor and measuring point memory)
5
Plug of the exciting current cable
6
Cable connector for power supply/pulse output/status output
7
Phillips screw
8
Electronics module
9
Socket for device connection
3
4
5
6
2
3
5
4
6
8
1
7
9
9