Technical data
Dosimass
34
Hauser
10.1.4
Output
Output signal
Pulse output:
Passive, max. 30 V DC/25 mA, pulse value and pulse polarity can be selected,
pulse width adjustable (0.05 ms to 1 s).
!
Note!
The device may only be connected to SELV, PELV or CLASS 2 circuits.
Signal on alarm
Pulse output
behavior can be selected
Transistor status output not conducting in the event of a fault/notice (depending on setting) or if
the power supply fails
Switching output
Status output:
Passive, max. 30 VDC/25 mA
!
Note!
The device may only be connected to SELV, PELV or CLASS 2 circuits.
Low flow cutoff
Switch point for low flow cutoff selectable.
Galvanic isolation
The power supply and outputs are galvanically isolated from one another.
10.1.5
Power supply
Electrical connections
12 ff.
Supply voltage
24 VDC nominal voltage (20 to 30 VDC)
!
Note!
• The power supply may not exceed a maximum short-circuit current of 50 A.
• The device may only be connected to SELV, PELV or CLASS 2 circuits.
Potential equalization
No special measures are necessary for potential equalization. For devices for the Ex area, see the
notes in the Ex-specific supplement to these Operating Instructions.
Cable connection
Lumberg plug (RSE8, M12 × 1) for power supply and signal outputs
Cable specification
Every suitable cable with a temperature specification at least 20 °C (68 °F) higher than the ambient
temperature in the application. We recommend you use a cable with a temperature specification of
+80 °C (+176 °F).
Power consumption
Max. 4.3 W
Switch-on current: max. 1A (< 6 ms)
Power supply failure
Lasting min. 20 ms:
All sensor and measuring point data remain in the DAT
Содержание dosimass