Troubleshooting
Dosimass
26
Hauser
9.4
System error messages (FieldCare)
Serious system errors are always recognized by the device as "Fault messages" (SF = System Fault
message) and displayed appropriately in the FieldCare operating program.
On the other hand, simulations are only classed and displayed as notice messages (SN = System
Notice message).
"
Caution!
In the event of a serious fault, a flowmeter might have to be returned to the manufacturer for repair.
The procedures on
5 must be carried out before you return the device to Hauser.
Always enclose a duly completed "Declaration of contamination" form. You will find a preprinted
form at the back of this manual.
Type No./error message
Cause
Remedy / spare part
S = System error
F = Fault message (
with
an impact on the pulse output)
N = Notice (
without
an impact on the pulse output)
System error - fault messages (LED = red, flashes quickly)
SF
# 011
AMP HW EEPROM
Electronics module:
Defective EEPROM
Replace electronics module
30.
SF
# 012
AMP SW EEPROM
Electronics module:
Error accessing EEPROM data
The EEPROM data blocks in which an error
has occurred are displayed in the RESTORE
DATA FAILURE function.
Faulty parameters are then replaced by
predefined default values.
SF
# 031
HW DAT
Sensor DAT:
1.
DAT is defective.
2.
DAT is not plugged in or is
missing.
1.
Replace DAT.
Spare parts:
Check the spare part set number to
ensure that the new, replacement DAT
is compatible with the measuring
electronics.
2.
Insert DAT:
SF
# 032
SW DAT
Sensor:
Error accessing the calibration values
stored in the DAT.
1.
Check whether the DAT is correctly
plugged in
30.
2.
Replace DAT. Spare parts
Before replacing the DAT, check that
the new, replacement DAT is
compatible with the measuring
electronics. Check the:
– Spare part set number
– Hardware revision code
3.
Replace the electronics module if
necessary
30.
Spare parts:
SF
# 359
PULSE RANGE
Pulse output:
Pulse output frequency is out of range.
1.
Increase the setting for pulse value.
2.
Reduce flow
SF
# 379
LOWER
FREQUENCY-LIMIT
The measuring tube oscillation
frequency is below the permitted
range.
Causes:
– Damaged measuring tube
– Sensor defective or damaged
Contact your Hauser service
organization.
SF
# 380
UPPER
FREQUENCY-LIMIT
The measuring tube oscillation
frequency is above the permitted
range.
Causes:
– Damaged measuring tube
– Sensor defective or damaged
Contact your Hauser service
organization.