Appendix - Function description
Hauser
47
11.5
Function group COMMUNICATION
11.6
Function group PROCESSPARAMETER
ACTUAL STATUS
Use this function to display the current status of the status output.
DISPLAY:
NON CONDUCTIVE
CONDUCTIVE
Function description
Function group COMMUNICATION
TAG NAME
Use this function to assign a tag name to the measuring device.
User input:
Max. 8-character text
Factory setting:
"_ _ _ _ _ _ _ _" (without text)
Function description
Function group PROCESS PARAMETER
ASSIGN LOW-FLOW
CUTOFF
Use this function to assign the measured variable for the low flow cutoff.
Options
:
MASS FLOW
VOLUME FLOW
Factory setting
:
MASS FLOW
ON VALUE LOW-FLOW
CUTOFF
For entering the on-value for low flow cutoff. Low flow cutoff is active if the setting is
a value not equal to 0. Low flow cutoff works with a hysteresis of 50% (off-value = 150%
of the on-value).
User input:
Multi-digit floating-point number
Factory setting:
The following factory settings correspond to a flow velocity of approx. v = 0.04 m/s.
SI units:
DN 8
8.00 kg/h
DN 15
26.00 kg/h
DN 25
72.00 kg/h
US units:
DN 8
0.300 lb/min
DN 15
1.000 lb/min
DN 25
2.600 lb/min
Function description
Function group STATUS OUTPUT